Selasa, 11 Maret 2014

MAMMOGRAPHY

M-Tex 30

Full Field Digital Mammography System

·       Dual LCD screen on column
·       Advanced flat panel technology a- Se or CMOS
·       Small tube head design
·       Isocentric rotation
·       Convenient Single- Touch Positioning
·       Convenient Single- Touch switch on/off
·       Large rotation range of tube head
·       Strong and user-friendly workstation
·       Unique design of compression device maximizing the comfortability of patient
·       Available for future upgrading with biopsy and 3D Breast Tomosynthesis


Technical Data M-tex 30
Sternmed Digital Mammography System
Main unit

Size
1110 (L) x650 (W) x2050 (H) mm
Weight
< 320Kg
C-arm rotating range
650~1400mm
C-arm rotating angle
+185°
SID
66CM
Dual-function pedals
2 pay
Patient protective mask
Removable
C-arm rotation angle, position,
oppressive force, oppression thickness be displayed digitally
C-arm, two displays on the column
Lifting range for compression plate
0-270mm (pressure plate to oppression platform height)(manual, electric)
Compression plate
Conventional compression plate: 24cm x 30cm
Conventional compression plate: 18 cm x 24 cm
9cm conventional point compression plate 9cm x 9cm
Enlarge oppression plate 18cm x 24cm
Enlarged compression plate 9cm x 9cm
Auxiliary compression plate
Small breasts oppression plate
Amplification system
1.8X magnification platform
1.5X magnification platform
Cabinet
Column dimensions
(length X width X height)
282mmx633mmx830mm
Weight
≤50Kg
Console (Optional)
Column dimensions
(length X width X height)
784mmX411.5mmX1071mm
Weight
≤30Kg
Tube: IAE
Ingredient
Molybdenum Target
Tube voltage range
20~40kV
Focus
0.1/0.3mm
Power
4.9kW
Anode rate
3000/10000rpm(Low / High)
Maximum anode dissipation rate
750W
Heat capacity
225kJ(300kHU)
Pipe sleeve of thermal
capacity
320kJ
Tube sets of maximum cooling rate
210W
Inherent filtration
0.5mm Be
Angle
10°/16°
Length
320mm
Maximum diameter
140mm
Weight
15kg
Cooling
Nature / air


High Voltage Generator: Spellman
Line Power 1
220V±20VAC, single phase, 50/60Hz
Line Power 2
380V±38VAC, three phase, 50/60Hz
Photography voltage
20-40KV (step 0.5KV)
Photography current
10-160mA (sub-file adjustment)
Maximum power shooting conditions (@ 0.1S)
125 mA @ 40 kVp
Nominal power
5kW
Photography Time
5ms-5s
Operating Frequency
25kHz
Maximum KV
40KV
mAs Range (mAs)
0.5mAS-500mAS
Control grids
There are (vibration grids)
AEC
There are (automatic AEC, semiautomatic AEC)
Collimator: Ralco
Maximum X-ray region
24X30cm(SID=66cm)
Light time
<15 s
Average illuminance
≥160LUX
Filtration
0.03mmMo,0.025mmRh
Adjustment method
Automatic
Power supply
24V DC, 2A
Weight
3.6Kg
Size
240 X175 X82MM
The shape of the light field
≤24X30cm (supports continuous adjustment)
HV Cable:
Power
60kVDC
DC voltage
90kVDC/15min
AC voltage
70kV/5min
Length
10m
FPD Detector: Anrad
Conversion Screen Type
Amorphous selenium
Effective pixel region
24cmХ30cm
Pixel matrix
2816X3584
Pixel size
85um
Spatial resolution limit
8lp/mm
Energy range
20~40kVp
A / D converter
16-bits
Acquisition cycle
8s (including image processing time was 10s)
Data transfer mode
Gigabit network transmission
Weight
3kg
Computer System
Monitor
LCD monitors, 19 inch, 1M color
Size
440mm*175mm*413mm
Volume
25L
Weight
8.5 kg
Chipset
INTER Q77
Processor
I5-3470 (VPRO) (CORE I5 3.2G 6M cache)
Integrated graphics
Integrated graphics
Sound Card
Integrated High Definition Audio
NIC
Motherboard integrated Gigabit Ethernet




Independent card
PRO/1000 PT Desktop Adapter
Memory Specifications
DDR3 1333
Memory
8G DDR3
Hard Specifications
SATA 7.2K
Hard drive capacity
500G hard disk partitions: C disk 60GB NTFS format, E disk 20GB FAT32 format, the remaining space forthe D drive NTFS format
COM interface
COM1
Drive bays
Interior: 2 * 5.25 1 * 3.5 external: 1 * 3.5
(expandable to 2 * 3.5)
Expansion Slots
2*PCI 2.3                 1*PCIeX1               1*PCIeX16
Software
Patient Management
patient information registration, save, set an appointment, query and other functions
Inspection management
parameter setting, exposure collection, automatic image processing
Image Browser
image selection, image processing (crop, zoom, grayscale adjustment, rotate, flip, tag, window layout, etc.)
Image post-processing
image enhancement, tissue equalization, noise reduction, gradation values of tensile
Image output
support for laser printer output, support for the laser camera output, DICOM interface and hard
Data Management
data deletion, statistics, edit, backup and recovery
Other features
detector calibration procedures, failure error function, Work list
module
Packing Information
packing size (mm)(L*W*H)
Weight 320 KG
Column : 2230*1060*1480
60
Cabinet: 800*490*1035
20
Accessories :1080*635*730
30
Compressing devices:
1080*635*730
30
Environment requirement
Work Environment
Temp:12°C~37°C
Humidity :20%~90% (No condensation)
Atmospheric pressure :70kPa~106kPa
Storage Environment
Temp:5°C~40°C
Humidity :20%~90%
Atmospheric pressure:70kPa~106kPa

PT. MEGA MEDIKA JAYA
Office :
22nd  Floor, Prudential Center, Kota Kasablanka
Jl. Casablanca Raya Kav. 88 JAKARTA 12870-INDONESIA
                                               
Operational Office :
Jl. Tanjung Barat Selatan No. 115 RT 01 RW02     
Kel. Lenteng Agung, Kec. Jagakarsa                        
JAKARTA SELATAN 12610  INDONESIA   

Phone  : 62-21-78830461       
Fax       : 62-21-78830461       
Mobile : 085770226999, 081213786999