Selasa, 11 Maret 2014

MRI

Marcom 1.5T

Superconductive MRI System

  • Main field strength: 1.5T
  • Spectrometer: Full Digital
  • Receiving Channel: 8
  • Receiving Coil: Phase Array Coils
  • Comprehensive scanning sequences
  • Advanced imaging techniques and clinical application
  • Very low helium consumption

      Technical data Marcom 1.5T
      Sternmed Superconductive 1.5T MRI
      Superconductive Magnet

      Superconductive Magnet
      1.5T active shield superconducting magnet
      Homogeneity
      ≤5.0 ppm @ DSV 45cm
      ≤1.0 ppm @ DSV 40cm
      Stability
      ≤0.02 ppm/h
      Helium boil-off rate
      <50 cc/h
      Helium refill period
      ≥24 months
      5GS (X,Y,Z)
      ≤2.5m,2.5m,4m
      Patient aperture
      600mm
      Gradient

      Gradient System
      Active shield full body gradient coil systems, noise reduction & mute technology
      Gradient Coil
      England Telsa Corporation
      Cooling System Type
      Water-cooled
      Cold Head
      Sumitomo cold head
      Gradient strength
      Single axis 34mT/m,  Effective 59 mT/m
      Slew rate
      Single axis 130mT/m/ms,  Effective 225mT/m/ms
      RF System

      Spectrometer
      All-digital (8 channnels)
      Power of transmitter amplifier
      18KW
      Receiving coils
      Built-in preamplifier phased array coil
      Receiving coil type (standard)
      head, neck, body, knee, spine, shoulders, Breast coil
      Workstation

      Operating system
      Windows XP Professional
      CPU
      Dual frequency ≥2.8GHz
      RAM
      ≥2GB
      Hard disk
      ≥300GB
      The main screen displays
      ≥22´´LCD TFT monitor
      Screen monitoring device, machine monitoring, patient monitoring, multi-parameter monitor the amount of concentrated, high security
      Machine control parameters
      pressure, level, temperature
      Patient monitoring parameters
      electrocardio, respiratory, RF, db/dt
      System operating modes
      normal operation mode, a controlled operating mode
      Image processing and analysis
      image enhancement, zoom, pan, crop, negatives, window width and window level adjustments, mark or clear text, distance measurement, regions of interest selected, the average pixel value, standard deviation of pixel values, signal values distribution, projection reconstruction
      Network components
      DICOM 3.0 standard interface, through the local Ethernet network easily to link camera, diagnosis and treatment workstations, medical information systems, remote diagnostics system.
      Pulse sequences


      2D, 3D GRE
      2D, 3D SE
      2D, 3D FSE
      FSPGR(positive/ negative  phase sequence)
      TrueFISP(Fast steady precession fast imaging)
      SSPGRE(Steady state process gradient echo)
      FRFSE(Fast recovery fast spin echo)
      Single-shot FSE 2D (Single shot fast spin echo)
      MSFSE(Multi shot fast spin echo)
      FSE 2D with Inversion Recovery (2D reversion fast spin echo sequence)
      STIR(Short time inversion recovery)
      FLAIR(Fluid attenuated inversion recovery)
      MRM,MRU,MRCP
      EPI SE-based
      EPI SE-based DW
      DWI(Diffusion weighted imaging)
      2D, 3D MRA
      Automatic pre-scan
      Gate Trigger
      Breathing, electrocardio
      Reconstruction algorithm
      1DFFT, 2DFFD, 2DFFT, MIP, MPR, SSD
      Scanning parameter

      Resolution
      0.75mm, Head,24cmFOV 256x256                                
      1.0mm, Body,  30cmFOV 256x256                                                                                                         0.5mm, Head , 24cmFOV 512x512
      0.75mm, Body,30cmFOV 512x512
      Acquisition matrix
      64/128/ 256/ 512/1024
      Maximum display matrix
      Maximum display matrix1024x1024
      FOV
      10~450 mm
      Scan orientations
      Any angle (axial, sagittal, coronal, any slope, multi-layer multi-angle)
      Image type
      T1 weighted imaging, T2 weighted imaging, T2*weighted imaging, proton density imaging, Water suppressed imaging, Fat Suppressed imagine, MRM, MRU, MRCP, Magnetic Resonance angiography(MRA), Diffusion weighted imaging(DWI)
      Patient Table

      Patient Table
      Drop out of the open two-dimensional movement, motor drives, cross laser positioning, Emergency braking situation or power outage, you can manually take the bed.
      Status display and Configuration
      8” TFT- LCD display, resolution 800 * 600, real-time display system status
      Max. Patient Load
      200Kg
      Positioning accuracy
      <1mm
      Operating position
      Bilateral
      Positioning Accessories
      mattress, pillow, head pillow, various parts of the fixed pad
      Power supply

      Helium compressor
      3N-a.c 380~415V/50Hz,6.9KW
      Machine
      3N-a.c.380V±38V/50Hz±1Hz,50KW q
      Typical Layout

      Magnet Room
      About 33m2(5m×6.5m);
      Equipment Room
      About 24m2(3m×8m);
      Control Room
      About 7.5m2(1.5m×5m);
      Total MRI-System area
      About 64.5m2.

PT. MEGA MEDIKA JAYA
Office :
22nd  Floor, Prudential Center, Kota Kasablanka
Jl. Casablanca Raya Kav. 88 JAKARTA 12870-INDONESIA
                                               
Operational Office :
Jl. Tanjung Barat Selatan No. 115 RT 01 RW02     
Kel. Lenteng Agung, Kec. Jagakarsa                        
JAKARTA SELATAN 12610  INDONESIA   

Phone  : 62-21-78830461       
Fax       : 62-21-78830461       
Mobile : 085770226999, 081213786999